Thermo Scientific Alpha pH 800 pH/ORP Controllers/Transmitters


Model: Alpha pH800
Hãng: Eutech
Specifications
Product Specification
Description
pH
Range-2.00 to 16.00 pH
Resolution0.01 pH
Accuracy±0.01 pH
ORP 
Range-1000 to 1000 mV
Resolution1 mV
Accuracy±1 mV
Temperature
Range-9.9 to 125.0 ºC
Resolution0.1 ºC
Accuracy±0.5 ºC
SensorPt100 / Pt1000 (jumper selectable); 2 or 3 wire
CompensationAuto/manual
Set Point And Controller Functions
Set point 1 (SP1) /
set point 2 (SP2)
-2.00 to 16.00 pH or -1000 to 1000 mV
Switching pH Hysteresis0.1 to 1 pH
Switching ORP Hysteresis10 to 100 mV
Function (Switchable)Limit control
Pickup/Dropout Delay0 to 2000 sec
Contact Outputs2 SPDT relays
Switching Voltage/
Current/Power
Max. 250 VAC / max. 3 A / max. 600 VA
Electrical Data And Connections
Transmitter Function0/4 to 20 mA scalable outputs for pH/ORP, galvanically isolated
Hold Function SwitchTo freeze output current and deactivate control relays
LoadMax. 600 Ohm
pH/ORP InputBNC (1012 impedance); asymmetrical/symmetrical
Connection Terminal5-pole, 17-pole terminal, detachable blocks
Display
LCD7 segments display with symbols for status information
Power Supply
Input110 VAC
(jumper selectable) ;
48 to 62 Hz ;
max. 0.75 VA
(For TSCTP0801)
220 VAC
(jumper selectable) ;
48 to 62 Hz ;
max. 0.75 VA
(For TSCTP0802)
Main FuseSlow-blow 250 V / 100 mA
Pollution Degree2
Transient Overvoltage CategoryII
EMC Specifications
Emitted InterferenceAccording to EN 50081-1
Immunity To InterferenceAccording to EN 50082-1
Environmental Conditions
Operating Temperature Range-10 to 50 ºC
Max. Relative Humidity80 % up to 31 ºC decreasing linearly to 50 %
at 40 ºC
Mechanical Specifications
Dimensions (WxHxD)96 x 96 x 175 mm
Weight700 g (unit) / 800 g (packed)
Ingress ProtectionIP54 (front panel)